| |
 |
COMMERCIAL TUBE
PRODUCT SELECTOR |
| |
|
|
 |
 |
| |
 |
MILITARY TUBE
PRODUCT SELECTOR |
| |
|
|
EPA Method TO-14
Compendium of Methods for the Determination of Toxic Organic Compounds
in Ambient Air - Compendium Method TO-14A Determination Of Volatile
Organic Compounds (VOCs) In Ambient Air Using Specially Prepared
Canisters With Subsequent Analysis By Gas Chromatography
METHOD TO-14
Compendium Method TO-14A Determination Of Volatile Organic Compounds
(VOCs) In Ambient Air Using Specially Prepared Canisters With Subsequent
Analysis By Gas Chromatography
1. Scope
1.1 This document describes a procedure for sampling and analysis
of volatile organic compounds (VOCs) in ambient air. The method was
originally based on collection of whole air samples in SUMMA® passivated
stainless steel canisters, but has now been generalized to other
specially prepared canisters (see Section 7.1.1.2). The VOCs are
separated by gas chromatography and measured by a mass spectrometer
or by multidetector techniques. This method presents procedures for
sampling into canisters to final pressures both above and below atmospheric
pressure (respectively referred to as pressurized and subatmospheric
pressure sampling).
1.2 This method is applicable to specific VOCs that have been tested
and determined to be stable when stored in pressurized and sub-atmospheric
pressure canisters. Numerous compounds, many of which are chlorinated
VOCs, have been successfully tested for storage stability in pressurized
canisters (1-3). However, minimal documentation is currently available
demonstrating stability of VOCs in subatmospheric pressure canisters.
1.3 The Compendium Method TO-14A target list is shown in Table
1. These compounds have been successfully stored in canisters and
measured at the parts per billion by volume (ppbv) level. This method
applies under most conditions encountered in sampling of ambient
air into canisters. However, the composition of a gas mixture in
a canister, under unique or unusual conditions, will change so that
the sample is known not to be a true representation of the ambient
air from which it was taken. For example, low humidity conditions
in the sample may lead to losses of certain VOCs on the canister
walls, losses that would not happen if the humidity were higher.
If the canister is pressurized, then condensation of water from high
humidity samples may cause fractional losses of water-soluble compounds.
Since the canister surface area is limited, all gases are in competition
for the available active sites. Hence an absolute storage stability
cannot be assigned to a specific gas. Fortunately, under conditions
of normal usage for sampling ambient air, most VOCs can be recovered
from canisters near their original concentrations after storage times
of up to thirty days.
2. Summary of Method
2.1 Both subatmospheric pressure and pressurized sampling modes
typically use an initially evacuated canister and pump-ventilated
sample line during sample collection. Pressurized sampling requires
an additional pump to provide positive pressure to the sample canister.
A sample of ambient air is drawn through a sampling train comprised
of components that regulate the rate and duration of sampling into
a pre-evacuated specially prepared passivated canister.
2.2 After the air sample is collected, the canister valve is closed,
an identification tag is attached to the canister, a chain-of-custody
(COC) form completed, and the canister is transported to a predetermined
laboratory for analysis.
2.3 Upon receipt at the laboratory, the canister tag data is recorded,
the COC completed, and the canister is attached to the analytical
system. During analysis, water vapor is reduced in the gas stream
by a Nafion® dryer (if applicable), and the VOCs are then concentrated
by collection in a cryogenically-cooled trap. The cryogen is then
removed and the temperature of the trap is raised. The VOCs originally
collected in the trap are revolatilized, separated on a GC column,
then detected by one or more detectors for identification and quantitation.
2.4 The analytical strategy for Compendium Method TO-14A involves
using a high-resolution gas chromatograph (GC) coupled to one or
more appropriate GC detectors. Historically, detectors for a GC have
been divided into two groups: non-specific detectors and specific
detectors. The non-specific detectors include, but are not limited
to, the nitrogen-phosphorus detector (NPD), the flame ionization
detector (FID), the electron capture detector (ECD) and the photo-ionization
detector (PID). The specific detectors include the linear quadrupole
mass spectrometer (MS) operating in either the select ion monitoring
(SIM) mode or the SCAN mode,
or the ion trap detector (see Compendium Method TO-15). The use of
these detectors or a combination of these detectors as part of the
analytical scheme is determined by the required specificity and sensitivity
of the application. While the non-specific detectors are less expensive
per analysis and in some cases far more sensitive than the specific
detectors, they vary in specificity and sensitivity for a specific
class of compounds. For instance, if multiple halogenated compounds
are targeted, an ECD is usually chosen; if only compounds containing
nitrogen or phosphorus are of interest, a NPD can be used; or, if
a variety of hydrocarbon compounds are sought, the broad response
of the FID or PID is appropriate. In each of these cases, however,
the specific identification of the compound within the class is determined
only by its retention time, which can be subject to shifts or to
interference from other non-targeted compounds. When misidentification
occurs, the error is generally a result of a cluttered chromatogram,
making peak assignment difficult. In particular, the more volatile
organics (chloroethanes, ethyltoluenes, dichlorobenzenes, and various
freons) exhibit less well defined chromatographic peaks, leading
to possible misidentification when using nonspecific detectors. Quantitative
comparisons indicate that the FID is more subject to error than the
ECD because the ECD is a much more selective detector and exhibits
a stronger response. Identification errors, however, can be reduced
by: (a) employing simultaneous detection by different detectors or
(b) correlating retention times from different GC columns for confirmation.
In either case, interferences on the non-specific detectors can still
cause error in identifying compounds of a complex sample. The non-specific
detector system (GC/NPD/FID/ECD/PID), however, has been used for
approximate quantitation of relatively clean samples. The non-specific
detector system can provide a "snapshot"
of the constituents in the sample, allowing determination of:
— Extent of misidentification due to overlapping peaks.
— Determination of whether VOCs are within or not within concentration
range, thus requiring further analysis by specific detectors (GC/MS/SCAN/SIM)
(i.e., if too concentrated, the sample is further diluted).
— Provide data as to the existence of unexpected peaks which require
identification by specific detectors.
On the other hand, the use of specific detectors (MS coupled to
a GC) allows positive compound identification, thus lending itself
to more specificity than the multidetector GC. Operating in the SIM
mode, the MS can readily approach the same sensitivity as the multidetector
system, but its flexibility is limited. For SIM operation the MS
is programmed to acquire data for a limited number of targeted compounds.
In the SCAN mode, however, the MS becomes a universal detector, often
detecting compounds which are not detected by the multidetector approach.
The GS/MS/SCAN will provide positive identification, while the GC/MS/SIM
procedure provides quantitation of a restricted list of VOCs, on
a preselected target compound list (TCL). If the MS is based upon
a standard ion trap design, only a scanning mode is used (note however,
that the Select Ion Storage (SIS) mode of the ion trap has features
of the SIM mode). See Compendium Method TO-15 for further explanation
and applicability of the ion-trap to the analysis of VOCs from specially
prepared canisters. The analyst often must decide whether to use
specific or non-specific detectors by considering such factors as
project objectives, desired detection limits, equipment availability,
cost and personnel capability in developing an analytic strategy.
A list of some of the advantages and disadvantages associated with
non-specific and specific detectors may assist the analyst in the
decision-making process.
Non-specific Multidetector Analytical System
Advantages Disadvantages
Advantages
! Somewhat lower equipment cost than GC/MS
! Less sample volume required for analysis
! More sensitive
- ECD may be 1000 times more sensitive than GC/MS
Disadvantages
! Multiple detectors to calibrate
! Compound identification not positive
! Lengthy data interpretation (1 hour each for analysis and data
reduction)
! Interference(s) from co-eluting compound(s)
! Cannot identify unknown compounds
- outside range of calibration
- without standards
! Does not differentiate targeted compounds from interfering compounds Specific Detector Analytical System
GC/MS/SIM
Advantages
! positive compound identification
! greater sensitivity than GC/MS/SCAN
! less operator interpretation than for multidetector GC
! can resolve co-eluting peaks
! more specific than the multidetector GC
Disadvantages
! cannot identify nonspecified compounds (ions)
! somewhat greater equipment cost than multidetector GC
! greater sample volume required than for multidetector GC
! universality of detector sacrified to achieve enhancement in sensitivity GC/MS/SCAN
Advantages
! positive compound identification
! can identify all compounds
! less operator interpretation
! can resolve co-eluting peaks
Disadvantages
! lower sensitivity than GC/MS/SIM
! greater sample volume required than for multidetector GC
! somewhat greater equipment cost than multidetector GC
The analytical finish for the measurement chosen by the analyst
should provide a definitive identification and a precise quantitation
of volatile organics. In a large part, the actual approach to these
two objectives is subject to equipment availability. Figure 1 indicates
some of the favorite options that are used in Compendium Method TO-14A.
The GC/MS/SCAN option uses a capillary column GC coupled to a MS
operated in a scanning mode and supported by spectral library search
routines. This option offers the nearest approximation to unambiguous
identification and covers a wide range of compounds as defined by
the completeness of the spectral library. GC/MS/SIM mode is limited
to a set of target compounds which are user defined and is more sensitive
than GC/MS/SCAN by virtue of the longer dwell times at the restricted
number of m/z values. Both these techniques, but especially the GC/MS/SIM
option, can use a supplemental general nonspecific detector to verify/identify
the presence of VOCs. Finally the option labelled GC-multidetector
system uses a combination of retention time and multiple general
detector verification to identify compounds. However, interference
due to nearly identical retention times can affect system quantitation
when using this option. Due to low concentrations of toxic VOCs encountered
in urban air (typically less than 25 ppbv and the majority below
10 ppbv) along with their complicated chromatographs, Compendium
Method TO-14A strongly recommends the specific detectors (GC/MS/SCAN/SIM)
for positive identification and for primary quantitation to ensure
that high-quality ambient data is acquired. For the experienced analyst
whose analytical system is limited to the non-specific detectors,
Section 10.3 does provide guidelines and example chromatograms showing
typical retention times and calibration response factors, and utilizing
the nonspecific detectors (GC/FID/ECD/PID) analytical system as the
primary quantitative technique. Compendium Method TO-15 is now available
as a guidance document containing additional advice on the monitoring
of VOCs. Method TO-15 contains information on alternative water management
systems, has a more complete quality control section, shows performance
criteria that any monitoring technique must achieve for acceptance,
and provides guidance specifically directed at compound identification
by mass spectrometry.
3. Significance
3.1 The availability of reliable, accurate and precise monitoring
methods for toxic VOCs is a primary need for state and local agencies
addressing daily monitoring requirements related to odor complaints,
fugitive emissions, and trend monitoring. VOCs enter the atmosphere
from a variety of sources, including petroleum refineries, synthetic
organic chemical plants, natural gas processing plants, biogenic
sources, and automobile exhaust. Many of these VOCs are toxic so
that their determination in ambient air is necessary to assess human
health impacts.
3.2 The canister-based monitoring method for VOCs has proven to
be a viable and widely used approach that is based on research and
evaluation performed since the early 1980s. This activity has involved
the testing of sample stability of VOCs in canisters and the design
of time-integrative samplers. the development of procedures for analysis
of samples in canisters, including the procedure for VOC preconcentration
from whole air, the treatment of water vapor in the sample, and the
selection of an appropriate analytical finish has been accomplished.
The canister-based method was initially summarized by EPA as Method
TO-14 in the First Supplement to the Compendium of Methods for the
Determination of Toxic Organic Compounds in Ambient Air. The present
document updates the original Compendium Method TO-14 with correction
of time-sensitive information and other minor changes as deemed appropriate.
3.3 The canister-based method is now a widely used alternative
to the solid sorbent-based methods. The method has sub-ppbv detection
limits for samples of typically 300-500 mL of whole air and duplicate
and replicate precisions under 20 percent as determined in field
tests. Audit bias values average within the range of ±10 percent.
These performance parameters are generally adequate for monitoring
at the 10-5 lifetime exposure risk levels for many VOCs.
3.4 Collection of ambient air samples in canisters provides a number
of advantages: (1) convenient integration of ambient samples over
a specific time period (e.g., 24 hours); (2) remote sampling and
central analysis; (3) ease of storing and shipping samples; (4) unattended
sample collection; (5) analysis of samples from multiple sites with
one analytical system; (6) collection of sufficient sample volume
to allow assessment of measurement precision and/or analysis of samples
by several analytical systems; and (7) storage stability for many
VOCs over periods of up to 30 days. To realize these advantages,
care must be exercised in selection, cleaning, and handling sample
canisters and sampling apparatus to avoid losses or contamination.
3.5 Interior surfaces of canisters are treated by any of a number
of passivation processes, one of which is SUMMA polishing as identified
in the original Compendium Method TO-14. Other specially prepared
canisters are also available (see Section 7.1.1.2).
3.6 The canister-based method for monitoring VOCs is the alternative
to the solid sorbent-based method described in conventional methods
such as the Compendium Methods TO-1 and TO-2, and in the new Compendium
Method TO-17 that describes the use multisorbent packings including
the use of new carbon-based sorbents. It also is an alternative to
on-site analysis in those cases where integrity of samples during
storage and transport has been established.
|