| |
 |
COMMERCIAL TUBE
PRODUCT SELECTOR |
| |
|
|
 |
 |
| |
 |
MILITARY TUBE
PRODUCT SELECTOR |
| |
|
|
EPA Method TO-15
Compendium of Methods for the Determination of Toxic Organic Compounds
in Ambient Air - Compendium Method TO-15 Determination Of Volatile
Organic Compounds (VOCs) In Air Collected In Specially-Prepared Canisters
And Analyzed By Gas Chromatography / Mass Spectrometry (GC/MS)
METHOD TO-15
Determination of Volatile Organic Compounds (VOCs) In Air Collected
In Specially-Prepared Canisters And Analyzed By Gas Chromatography/
Mass Spectrometry (GC/MS) 1. Scope
1.1 This method documents sampling and analytical procedures for
the measurement of subsets of the 97 volatile organic compounds (VOCs)
that are included in the 189 hazardous air pollutants (HAPs) listed
in Title III of the Clean Air Act Amendments of 1990. VOCs are defined
here as organic compounds having a vapor pressure greater than 10-1
Torr at 25EC and 760 mm Hg. Table 1 is the list of the target VOCs
along with their CAS number, boiling point, vapor pressure and an
indication of their membership in both the list of VOCs covered by
Compendium Method TO-14A (1) and the list of VOCs in EPA's Contract
Laboratory Program (CLP) document entitled: Statement-of-Work (SOW)
for the Analysis of Air Toxics from Superfund Sites (2).
Many of these compounds have been tested for stability in concentration
when stored in specially-prepared canisters (see Section 8) under
conditions typical of those encountered in routine ambient air analysis.
The stability of these compounds under all possible conditions is
not known. However, a model to predict compound losses due to physical
adsorption of VOCs on canister walls and to dissolution of VOCs in
water condensed in the canisters has been developed (3). Losses due
to physical adsorption require only the establishment of equilibrium
between the condensed and gas phases and are generally considered
short term losses, (i.e., losses occurring over minutes to hours).
Losses due to chemical reactions of the VOCs with cocollected ozone
or other gas phase species also account for some short term losses.
Chemical reactions between VOCs and substances inside the canister
are generally assumed to cause the gradual decrease of concentration
over time (i.e., long term losses over days to weeks). Loss mechanisms
such as aqueous hydrolysis and biological degradation (4) also exist.
No models are currently known to be available to estimate and characterize
all these potential losses, although a number of experimental observations
are referenced in Section 8. Some of the VOCs listed in Title III
have short atmospheric lifetimes and may not be present except near
sources.
1.2 This method applies to ambient concentrations of VOCs above
0.5 ppbv and typically requires VOC enrichment by concentrating up
to one liter of a sample volume. The VOC concentration range for
ambient air in many cases includes the concentration at which continuous
exposure over a lifetime is estimated to constitute a 10-6 or higher
lifetime risk of developing cancer in humans. Under circumstances
in which many hazardous VOCs are present at 10-6 risk concentrations,
the total risk may be significantly greater.
1.3 This method applies under most conditions encountered in sampling
of ambient air into canisters. However, the composition of a gas
mixture in a canister, under unique or unusual conditions, will change
so that the sample is known not to be a true representation of the
ambient air from which it was taken. For example, low humidity conditions
in the sample may lead to losses of certain VOCs on the canister
walls, losses that would not happen if the humidity were higher.
If the canister is pressurized, then condensation of water from high
humidity samples may cause fractional losses of water-soluble compounds.
Since the canister surface area is limited, all gases are in competition
for the available active sites. Hence an absolute storage stability
cannot be assigned to a specific gas. Fortunately, under conditions
of normal usage for sampling ambient air, most VOCs can be recovered
from canisters near their original concentrations after storage times
of up to thirty days (see Section 8).
1.4 Use of the Compendium Method TO-15 for many of the VOCs listed
in Table 1 is likely to present two difficulties: (1) what calibration
standard to use for establishing a basis for testing and quantitation,
and (2) how to obtain an audit standard. In certain cases a chemical
similarity exists between a thoroughly tested compound and others
on the Title III list. In this case, what works for one is likely
to work for the other in terms of making standards. However, this
is not always the case and some compound standards will be troublesome.
The reader is referred to the Section 9.2 on standards for guidance.
Calibration of compounds such as formaldehyde, diazomethane, and
many of the others represents a challenge.
1.5 Compendium Method TO-15 should be considered for use when a
subset of the 97 Title III VOCs constitute the target list. Typical
situations involve ambient air testing associated with the permitting
procedures for emission sources. In this case sampling and analysis
of VOCs is performed to determine the impact of dispersing source
emissions in the surrounding areas. Other important applications
are prevalence and trend monitoring for hazardous VOCs in urban areas
and risk assessments downwind of industrialized or source-impacted
areas.
1.6 Solid adsorbents can be used in lieu of canisters for sampling
of VOCs, provided the solid adsorbent packings, usually multisorbent
packings in metal or glass tubes, can meet the performance criteria
specified in Compendium Method TO-17 which specifically addresses
the use of multisorbent packings. The two sample collection techniques
are different but become the same upon movement of the sample from
the collection medium (canister or multisorbent tubes) onto the sample
concentrator. Sample collection directly from the atmosphere by automated
gas chromatographs can be used in lieu of collection in canisters
or on solid adsorbents.
2. Summary of Method
2.1 The atmosphere is sampled by introduction of air into a specially-prepared
stainless steel canister. Both subatmospheric pressure and pressurized
sampling modes use an initially evacuated canister. A pump ventilated
sampling line is used during sample collection with most commercially
available samplers. Pressurized sampling requires an additional pump
to provide positive pressure to the sample canister. A sample of
air is drawn through a sampling train comprised of components that
regulate the rate and duration of sampling into the pre-evacuated
and passivated canister.
2.2 After the air sample is collected, the canister valve is closed,
an identification tag is attached to the canister, and the canister
is transported to the laboratory for analysis.
2.3 Upon receipt at the laboratory, the canister tag data is recorded
and the canister is stored until analysis. Storage times of up to
thirty days have been demonstrated for many of the VOCs (5).
2.4 To analyze the sample, a known volume of sample is directed
from the canister through a solid multisorbent concentrator. A portion
of the water vapor in the sample breaks through the concentrator
during sampling, to a degree depending on the multisorbent composition,
duration of sampling, and other factors. Water content of the sample
can be further reduced by dry purging the concentrator with helium
while retaining target compounds. After the concentration and drying
steps are completed, the VOCs are thermally desorbed, entrained in
a carrier gas stream, and then focused in a small volume by trapping
on a reduced temperature trap or small volume multisorbent trap.
The sample is then released by thermal desorption and carried onto
a gas chromatographic column for separation. As a simple alternative
to the multisorbent/dry purge water management technique, the amount
of water vapor in the sample can be reduced below any threshold for
affecting the proper operation of the analytical system by reducing
the sample size. For example, a small sample can be concentrated
on a cold trap and released directly to the gas chromatographic column.
The reduction in sample volume may require an enhancement of detector
sensitivity. Other water management approaches are also acceptable
as long as their use does not compromise the attainment of the performance
criteria listed in Section 11. A listing of some commercial water
management systems is provided in Appendix A. One of the alternative
ways to dry the sample is to separate VOCs from condensate on a low
temperature trap by heating and purging the trap. 2.5 The analytical strategy for Compendium Method TO-15 involves
using a high resolution gas chromatograph (GC) coupled to a mass
spectrometer. If the mass spectrometer is a linear quadrupole system,
it is operated either by continuously scanning a wide range of mass
to charge ratios (SCAN mode) or by monitoring select ion monitoring
mode (SIM) of compounds on the target list. If the mass spectrometer
is based on a standard ion trap design, only a scanning mode is used
(note however, that the Selected Ion Storage (SIS) mode for the ion
trap has features of the SIM mode). Mass spectra for individual peaks
in the total ion chromatogram are examined with respect to the fragmentation
pattern of ions corresponding to various VOCs including the intensity
of primary and secondary ions. The fragmentation pattern is compared
with stored spectra taken under similar conditions, in order to identify
the compound. For any given compound, the intensity of the primary
fragment is compared with the system response to the primary fragment
for known amounts of the compound. This establishes the compound
concentration that exists in the sample. Mass spectrometry is considered
a more definitive identification technique than single specific detectors
such as flame ionization detector (FID), electron capture detector
(ECD), photoionization detector (PID), or a multidetector arrangement
of these (see discussion in Compendium Method TO-14A). The use of
both gas chromatographic retention time and the generally unique
mass fragmentation patterns reduce the chances for misidentification.
If the technique is supported by a comprehensive mass spectral database
and a knowledgeable operator, then the correct identification and
quantification of VOCs is further enhanced.
3. Significance
3.1 Compendium Method TO-15 is significant in that it extends the
Compendium Method TO-14A description for using canister-based sampling
and gas chromatographic analysis in the following ways:
• Compendium Method TO-15 incorporates a multisorbent/dry purge technique
or equivalent (see Appendix A) for water management thereby addressing
a more extensive set of compounds (the VOCs mentioned in Title III
of the CAAA of 1990) than addressed by Compendium Method TO-14A.
Compendium Method TO-14A approach to water management alters the
structure or reduces the sample stream concentration of some VOCs,
especially water-soluble VOCs.
• Compendium Method TO-15 uses the GC/MS technique as the only means
to identify and quantitate target compounds. The GC/MS approach provides
a more scientifically-defensible detection scheme which is generally
more desirable than the use of single or even multiple specific detectors.
• In addition, Compendium Method TO-15 establishes method performance
criteria for acceptance of data, allowing the use of alternate but
equivalent sampling and analytical equipment. There are several new
and viable commercial approaches for water management as noted in
Appendix A of this method on which to base a VOC monitoring technique
as well as other approaches to sampling (i.e., autoGCs and solid
adsorbents) that are often used. This method lists performance criteria
that these alternatives must meet to be acceptable alternatives for
monitoring ambient VOCs.
• Finally, Compendium Method TO-15 includes enhanced provisions for
inherent quality control. The method uses internal analytical standards
and frequent verification of analytical system performance to assure
control of the analytical system. This more formal and better documented
approach to quality control guarantees a higher percentage of good
data.
3.2 With these features, Compendium Method TO-15 is a more general
yet better defined method for VOCs than Compendium Method TO-14A.
As such, the method can be applied with a higher confidence to reduce
the uncertainty in risk assessments in environments where the hazardous
volatile gases listed in the Title III of the Clean Air Act Amendments
of 1990 are being monitored. An emphasis on risk assessments for
human health and effects on the ecology is a current goal for the
U.S. EPA.
|